首页> 外文会议>European Frequency and Time Forum >Absolute frequency measurement of the ytterbium ion E3 optical clock transition using international atomic time
【24h】

Absolute frequency measurement of the ytterbium ion E3 optical clock transition using international atomic time

机译:international离子E3光学时钟跃迁的绝对频率测量,采用国际原子钟

获取原文

摘要

We report a revised absolute frequency measurement of the 171Yb+electric octupole optical clock transition of 642 121 496 772 645.17(22) Hz, with a fractional uncertainty of 3.4 × 10-16. This newer value, with lower uncertainty, is achieved through additional analysis of previously reported experimental data that links the optical frequency to the SI second through international atomic time (TAI). The new step in the data analysis involves a special computation of the fractional frequency offset between the scale interval of TAI and the SI second, over a time interval more closely matched to the experimental data acquisition period.
机译:我们报告了171Yb的修订绝对频率测量结果 + 八极电时钟的电子跃迁为642121496496772645.17(22)Hz,不确定度为3.4×10-16。通过对以前报告的实验数据进行附加分析,可以实现这一具有较低不确定性的新值,该实验数据通过国际原子时间(TAI)将光频率与SI秒相联系。数据分析中的新步骤涉及在与实验数据采集周期更紧密匹配的时间间隔内,对TAI的标度间隔与SI秒之间的分数频率偏移进行特殊计算。