When an automatic exposure control is introduced in C-arm cone beam CT data acquisition, the spectral inconsistencies between acquired projection data are exacerbated. As a result, conventional water /bone correction schemes are not as effective as in conventional diagnostic x-ray CT acquisitions with a fixed tube potential. In this paper, a new method was proposed to reconstruct several images with different degrees of spectral consistency and thus different levels of beam hardening artifacts. The new method relies neither on prior knowledge of the x-ray beam spectrum nor on prior compositional information of the imaging object. Numerical simulations were used to validate the algorithm.
机译:当在C型臂锥束CT数据采集中引入自动曝光控制时,会加剧采集的投影数据之间的光谱不一致性。结果,常规的水/骨校正方案不如具有固定管电势的常规诊断X射线CT采集有效。在本文中,提出了一种新的方法来重建具有不同程度的光谱一致性并因此具有不同水平的光束硬化伪影的多个图像。新方法既不依赖于X射线束光谱的先验知识,也不依赖于成像对象的先验组成信息。数值模拟被用来验证算法。