首页> 外文会议>IEEE International Ultrasonics Symposium >Automatic measurement of the fetal abdominal section on a portable ultrasound machine for use in low and middle income countries
【24h】

Automatic measurement of the fetal abdominal section on a portable ultrasound machine for use in low and middle income countries

机译:在中低收入国家使用的便携式超声仪上自动测量胎儿腹部

获取原文

摘要

General pregnancy monitoring with ultrasound includes the measurement of the biparietal diameter (BPD), femur (FL) and the mean abdominal diameter (MAD) of the fetus. This study was aimed to develop an automatic method for localization of the presented section through the abdomen and measurement of MAD. The algorithm may be run on a conventional ultrasound machine or on a tablet-based machine designed to be used in the rural areas of low- and middle-income countries (LMIC). The open source computer vision (OpenCV) library was used to process a B-mode ultrasound image to detect the location of a fetal abdomen. A Kalman based tracker (RCTL, GE Vingmed Ultrasound) was used to converge a deformable circle model along the boundary of the detected abdomen. The mean of the model's major and minor axes was considered as MAD. The method automatically localized the abdomen and measured MAD in 57 of 61 images (93%) collected from 16 - 41-week-old fetuses. The correlation and the error plots between reference and automatic MAD measurements are presented. The correlation coefficient was 0.96, mean error was -0.06 mm and the error range (95% CI) was -14.80 to 14.68 mm. The resulting errors increased with the gestational age (GA) of the fetuses due to inner shadow and lack of edge information. The method was tested to perform well both in a PC and on various mobile devices.
机译:用超声波进行的一般妊娠监测包括胎儿双顶径(BPD),股骨(FL)和平均腹部直径(MAD)的测量。这项研究旨在开发一种通过腹部和MAD测量来定位所呈现部分的自动方法。该算法可以在常规的超声波机器上运行,也可以在为中低收入国家(LMIC)的农村地区设计的基于平板电脑的机器上运行。开源计算机视觉(OpenCV)库用于处理B型超声图像,以检测胎儿腹部的位置。使用基于卡尔曼的跟踪器(RCTL,GE Vingmed Ultrasound)沿检测到的腹部边界收敛可变形的圆形模型。模型的长轴和短轴的平均值被认为是MAD。该方法可自动定位腹部并测量从16-41周龄胎儿收集的61幅图像(占93%)中的57幅MAD。给出了参考和自动MAD测量之间的相关性和误差图。相关系数为0.96,平均误差为-0.06毫米,误差范围(95%CI)为-14.80至14.68毫米。由于内部阴影和缺乏边缘信息,导致的错误随着胎儿的胎龄(GA)而增加。经过测试,该方法在PC和各种移动设备上均具有良好的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号