首页> 外文会议>International Symposium on Antennas and Propagation >FEM simulation of induced interference voltage at implantable cardiac pacemaker due to wireless power transfer in HF-band
【24h】

FEM simulation of induced interference voltage at implantable cardiac pacemaker due to wireless power transfer in HF-band

机译:HF频段中无线功率传输引起的植入式心脏起搏器感应干扰电压的有限元模拟

获取原文
获取外文期刊封面目录资料

摘要

A novel numerical Wireless Power Transfer (WPT)/Active Implantable Medical Device-EMI estimation methodology based upon finite element method analysis is presented. This assessment methodology can be applied to both Mobile/Portable-WPT system and electric-vehicle-WPT system. Here, an example for HF-band magnetic resonance type WPT is introduced. Simulated example includes different exposure scenarios in order to estimate the maximum interference voltage.
机译:提出了一种基于有限元方法分析的新型数值无线功率传输(WPT)/有源植入式医疗器械-EMI估计方法。该评估方法可以应用于移动/便携式WPT系统和电动WPT系统。这里,介绍了用于HF频带磁共振型WPT的示例。仿真示例包括不同的暴露场景,以便估算最大干扰电压。