首页> 外文会议>IEEE MTT-S International Microwave Symposium >Adaptive harmonics comb notch digital filter for measuring heart rate of laboratory rat using a 60-GHz radar
【24h】

Adaptive harmonics comb notch digital filter for measuring heart rate of laboratory rat using a 60-GHz radar

机译:自适应谐波梳齿陷波数字滤波器,用于使用60 GHz雷达测量实验大鼠的心率

获取原文

摘要

Harmonics of respiration frequency due to nonlinear Doppler phase demodulation may interfere the identification of heartbeat frequency in vital sign radar's baseband spectrum. The proposed adaptive harmonics comb notch digital filter is a type of spectrum-based filtering function with no degradation on the heartbeat signal, and no calibration is needed when the detection distance changes. Experimental results show that the proposed filter for removing respiration harmonics can help measuring laboratory rat's heart rate more accurately when the heartbeat movement is weak or overwhelmed by the respiratory movement in the detected spectrum. Furthermore, even when the heart rate is totally overlapped by the respiration rate harmonics, the heart rate can still be extracted.
机译:非线性多普勒相位解调引起的呼吸频率谐波可能会干扰生命体征雷达基带频谱中心跳频率的识别。提出的自适应谐波梳状陷波数字滤波器是一种基于频谱的滤波功能,对心跳信号无衰减,并且当检测距离改变时无需校准。实验结果表明,在检测到的频谱中,当心跳运动微弱或被呼吸运动压倒时,所提出的消除呼吸谐波的滤波器可以帮助更准确地测量实验大鼠的心率。此外,即使当心率被呼吸率谐波完全重叠时,仍然可以提取心率。