首页> 外文会议>Conference on physics of medical imaging >Comparison of two CDMAM generations with respect to dose sensitivity
【24h】

Comparison of two CDMAM generations with respect to dose sensitivity

机译:两种CDMAM代在剂量敏感性方面的比较

获取原文

摘要

A contrast-detail phantom like the CDMAM phantom (Artinis Medical Systems, Zetten, NL) is suggested by the 'European protocol for the quality control of the physical and technical aspects of mammography screening' to evaluate image quality of digital mammography systems. In a recent paper the commonly used CDMAM 3.4 was evaluated according to its dose sensitivity in comparison to other phantoms. The successor phantom (CDMAM 4.0) features other disc diameters and thicknesses that were adapted to be more closely to the image quality which can be found in modern mammography systems. It seems to be obvious to compare this two generations of phantoms with respect to a potential improvement. The time-current product was varied within a range of clinically used values (40-160 mAs). Image evaluation was performed using the automatic evaluation software provided by Artinis. The relative dose sensitivity was compared in dependence of different diameters. Additionally, the IQFinv parameter, which averages over the diameters was computed to get a more global conclusion. We found that the dose is of a considerable smoother dependence with the CMDAM 4.0 phantom. Also the IQFinv parameter shows a more linear behaviour than with the CDMAM 3.4. As the automatic evaluation shows different results on the two phantoms, conversion factors from automatic to human readouts have to be adapted consequently.
机译:像CDMAM体模(Artinis Medical Systems,Zetten,NL)一样的对比细节体模由“用于乳腺X线检查筛查的物理和技术方面质量控制的欧洲协议”建议,以评估数字乳腺X线摄影系统的图像质量。在最近的一篇论文中,根据其与其他体模相比的剂量敏感性对常用的CDMAM 3.4进行了评估。后继体模(CDMAM 4.0)具有其他盘片直径和厚度,这些盘片的直径和厚度经过调整,使其更接近现代乳腺摄影系统中可以找到的图像质量。比较这两代幻影在潜在改进方面似乎是显而易见的。时间电流乘积在临床使用的值(40-160 mAs)范围内变化。使用Artinis提供的自动评估软件进行图像评估。根据不同的直径比较了相对剂量敏感性。此外,计算了平均直径的IQFinv参数以获得更全面的结论。我们发现该剂量对CMDAM 4.0体模具有相当平滑的依赖性。此外,IQFinv参数显示出比CDMAM 3.4更线性的行为。由于自动评估在两个体模上显示出不同的结果,因此必须调整从自动读数到人为读数的转换因子。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号