首页> 外文会议>IEEE International Conference on E-Health and Bioengineering >A new approach of shielding technique due to a leakage of radiation exposure outside a 10 MEV Linac room in a facility designed for 6 MEV
【24h】

A new approach of shielding technique due to a leakage of radiation exposure outside a 10 MEV Linac room in a facility designed for 6 MEV

机译:由于6 MEV设计的设施中的10 MEV直线加速器室之外的辐射暴露泄漏而产生的一种新的屏蔽技术

获取原文

摘要

At the national cancer institute - Cairo - Egypt. A decision of upgrading a linear accelerator of 6 MeV, Elekta machine, to high energy 10 MeV is taken. Because of advances in radiation sources and techniques, shielding and room design considerations for radiation oncology and such upgrade, are not covered by the NCRP reports. A problem has arises due to limited room size that was previously shielded based upon the 6 MeV energy, there was no other spaces for additional shielding in the treatment room which creates a problem of radiation leakage. This work presents an appraisal of radiation leakage and calculations of doses which vary from 65μSv/hr to 7.5 mSv/hr. We discuss the particular shielding problems related to high energy machines. Finally a new shielding methodology is presented merging more than two materials shows to be effective and reduce shielding cost.
机译:在国家癌症研究所-开罗-埃及。决定将Elekta机器的6 MeV线性加速器升级为高能量10 MeV。由于辐射源和技术的进步,NCRP报告未涵盖辐射肿瘤学和此类升级的屏蔽和房间设计注意事项。由于先前基于6 MeV能量屏蔽的有限的房间大小而引起了问题,在治疗室中没有其他空间可用于额外的屏蔽,这会产生辐射泄漏的问题。这项工作提出了对辐射泄漏的评估,并计算了从65μSv/ hr到7.5 mSv / hr的剂量。我们讨论与高能机器有关的特殊屏蔽问题。最后,提出了一种新的屏蔽方法,该方法将两种以上的材料融合在一起是有效的并降低了屏蔽成本。