We present a medical resonance imaging (MRI) B1-field polarisation mapping system based on active electro-optical H-field probes. The active electro-optical field probes have a bandwidth ranging from 10 MHz to 6 GHz with a sensitivity of <; -110dB(A/m)/√Hz at 128MHz. The miniature size of the probe elements (2×2 mm) provides a high spatial resolution. The electrical isolation using fiber-optics enables the system to be used in the harsh electro-magnetic environment of an MRI system and eliminates any feedback on the RF transmit coil under test. The developed measurement system has been tailored to the B1-field measurement application with a specific control and postprocessing software and allows for traceable B-field calibration with an absolute measurement uncertainty of better than ±1 dB. The relative measurement uncertainties in terms of the axial ratio and the relative phase of the B1x and B1y (polarisation ellipse parameter) are better than 0.4 dB and 3%, respectively.
机译:我们提出了基于有源电光H场探头的医学共振成像(MRI)B1场极化映射系统。有源电光探头的带宽范围为10 MHz至6 GHz,灵敏度为<;在128MHz时为-110dB(A / m)/√Hz。探针元件的微型尺寸(2×2 mm)提供了很高的空间分辨率。使用光纤的电隔离使该系统可以在MRI系统的恶劣电磁环境中使用,并消除了被测RF发射线圈上的任何反馈。开发的测量系统已通过特定的控制和后处理软件针对B1场测量应用进行了量身定制,并允许可溯源的B场校准,绝对测量不确定度优于±1 dB。 B1x和B1y(极化椭圆参数)的轴向比率和相对相位方面的相对测量不确定度分别优于0.4 dB和3%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号