首页> 外文会议>IEEE Nuclear Science Symposium and Medical Imaging Conference >The BC-704 scintillation screen with light readout by wavelength shifting fibers as a highly efficient neutron detector
【24h】

The BC-704 scintillation screen with light readout by wavelength shifting fibers as a highly efficient neutron detector

机译:BC-704闪烁筛网,具有光读出光的光读数,作为高效中子探测器

获取原文

摘要

The aim of the work is to investigate the revised concept of a neutron detector, offered by Saint Gobain-the BC-704 scintillation screen with a light readout by Wavelength Shifting (WLS) fibers. This detector was developed as a replacement of helium tubes, due to the lack of 3He on the world market. It is characterized by high sensitivity to gamma-rays, but the events from neutrons and gammas are well separated. Measurements with gamma and neutron sources have been performed and the results have been compared with helium-3 counter. According to American National Standards Institute (ANSI), we obtained n/γ discrimination on the level of 10−10, whereas absolute neutron detection efficiency was estimated as approximately two times higher than in the case of the reference 3He tube without the moderator. The tested detector is a sample prototype of one detector configuration of many possible, and that it can be modified to the neutron efficiency specification required.
机译:这项工作的目的是调查由圣甘油 - BC-704闪烁筛网提供的中子探测器的修订概念,通过波长移位(WLS)纤维,光读出光。由于世界市场缺乏3HE,该探测器是作为氦气管的替代品。它的特征在于对伽马射线的高敏感性,但是中子和伽马酸的事件分开很好。已经进行了用伽马和中子源进行测量,并将结果与​​氦-3计数器进行比较。据美国国家标准研究所(ANSI)表示,我们在10 -10 的水平上获得了n /γ歧视,而绝对中子检测效率估计大约比参考的情况高约两倍 3 他管没有主持人。测试检测器是一种探测器配置的样品原型,许多可能的配置,并且可以修改到所需的中子效率规格。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号