首页> 外文会议>Air quality measurement methods and technology conference >Field Testing of New Interference-Free O_3 Monitors
【24h】

Field Testing of New Interference-Free O_3 Monitors

机译:新型无干扰O_3监视器的现场测试

获取原文

摘要

The new humidity-controlled 2B 211 and TAPI 265E O_3 monitors appear virtually free of interferent bias compared to conventional FEM photometers. Field quality assurance audit and calibration protocols for conventional photometer compliance networks should be amended to include O_3 standard addition tests to ambient air to better assess local interference issues. Agency APEX O_3 exposure, risk, and regulatory benefit assessment ME factors and indoor penetration-deposition models should be corrected to account for the positive conventional photometer measurement bias that leads to overestimates of O_3 risk.
机译:与传统的FEM光度计相比,新型湿度控制2B 211和TAPI 265E O_3监视器几乎没有干扰偏差。常规光度计一致性网络的现场质量保证审核和校准协议应进行修改,以包括对环境空气的O_3标准附加测试,以更好地评估本地干扰问题。应该纠正代理商APEX O_3的暴露,风险和监管效益评估的ME因素和室内渗透沉积模型,以解决导致过度估计O_3风险的常规光度计测量偏差。