首页> 外文会议>IEEE International Instrumentation and Measurement Technology Conference >Highly sensitive microwave resonant near-field sensor for precise aqueous glucose detection in microfluidic medical applications
【24h】

Highly sensitive microwave resonant near-field sensor for precise aqueous glucose detection in microfluidic medical applications

机译:高敏感的微波共振近场传感器,用于微流体医疗应用中的精确含水葡萄糖检测

获取原文
获取外文期刊封面目录资料

摘要

This paper presents a highly sensitive resonant near-field sensor to detect small variations of glucose concentrations in water. The sensor structure is based on a planar half-wavelength (lambda/2) microstrip line (MSL) resonator with a microfluidic container placed on top. The interaction between the electric field and the milliliter range water-glucose solutions under test can be observed measuring the magnitude shift of the S-parameter from 1.96 to 2.04 GHz. Combining the microwave detection and the microfluidic network, a time efficient and low cost architecture is presented. With this technique, a very sensitive water-glucose concentration of less than 50 mg/dl can be separated.
机译:本文呈现出高度敏感的共振近场传感器,以检测水中葡萄糖浓度的小变化。传感器结构基于平面半波长(Lambda / 2)微带线(MSL)谐振器,其具有放置在顶部的微流体容器。可以观察到正在测试的电场和毫流范围水 - 葡萄糖溶液之间的相互作用测量从1.96到2.04 GHz的S参数的大小偏移。结合微波检测和微流体网络,提出了一种时间效率和低成本的架构。利用这种技术,可以分离非常敏感的水葡萄糖浓度小于50mg / dL。