摘要

We developed a highly sensitive waveguide-mode sensor of size 5 cm×10 cm×15 cm (Fig.1). The first topic discussed in this paper is that the discrimination of the H3N2 and H5N1 strains of influenza viruses has been clearly achieved with the waveguide-mode sensor. We found that glycan-based detection using the waveguide sensor system analysis is very useful to distinguish between H3N2 and H5N1 viruses. The second topic concerns sensitivity comparison among the waveguide-mode sensor, immuno-chromatography, enzyme-linked immunosorbent assay (ELISA), and surface plasmon resonance (SPR). The highest sensitivity for H1N1 and H3N2 viruses was found in ELISA and the waveguide-mode sensor.
机译:我们开发了尺寸为5 cm×10 cm×15 cm的高灵敏度波导模式传感器(图1)。本文讨论的第一个主题是使用波导模式传感器已经清楚地实现了对H3N2和H5N1流感病毒株的区分。我们发现使用波导传感器系统分析的基于聚糖的检测对于区分H3N2和H5N1病毒非常有用。第二个主题涉及波导模式传感器,免疫色谱,酶联免疫吸附测定(ELISA)和表面等离振子共振(SPR)之间的灵敏度比较。在ELISA和波导模式传感器中发现了对H1N1和H3N2病毒的最高灵敏度。