首页> 外文会议>IEEE International Symposium on Circuits and Systems >Performance evaluation of a low power optical wireless link for biomedical data transfer
【24h】

Performance evaluation of a low power optical wireless link for biomedical data transfer

机译:生物医学数据传输的低功耗光无线链路性能评估

获取原文

摘要

We report on an experimental low power transcutaneous optical telemetric link (TOTL) for high speed biomedical data transfer. By decreasing the bandwidth to data rate ratio, a silicon photodiode with large active size can be used to significantly increase transmission efficiency as well as misalignment tolerance. Additionally, a custom-designed transimpedance amplifier (TIA) with low input referred noise and a modified on-off keying with a custom designed low power VCSEL (vertical cavity surface emitting laser) driver are used to further reduce the required power consumption. The proposed system transmits data up to 75 Mbps through 6 mm thick tissue, in the presence of a 4 mm misalignment while achieving a BER better than 10−5, with a power consumption lower than 2.8 mW.
机译:我们报告了高速生物医学数据传输的实验低功率经皮光电遥测链路(TOTL)。通过将带宽降低到数据速率比,具有大的有源大小的硅光电二极管可用于显着提高传输效率以及未对准公差。另外,使用具有低输入噪声的定制设计的跨阻(TIA)和具有定制设计的低功率VCSEL(垂直腔表面发射激光)驱动器的改进的开关键控,用于进一步降低所需的功耗。所提出的系统在4mm未对准的情况下在4mm的未对准情况下将高达75 mbps的数据传输到6 mm厚的组织,同时实现比10-5更好,功耗低于2.8 mW。