首页> 外文会议>International Workshop on Magnetic Particle Imaging >First hybrid MPI-MRI imaging system as integrated design for mice and rats: Description of the instrumentation setup
【24h】

First hybrid MPI-MRI imaging system as integrated design for mice and rats: Description of the instrumentation setup

机译:首款用于小鼠和老鼠的一体化MPI-MRI混合成像系统:仪器设置说明

获取原文

摘要

This hybrid MPI-MRI imaging system (see Fig. 1 RIGHT) is currently under assembly at the research labs of Bruker BioSpin MRI GmbH, whereas single components were already tested. The magnetic field strength and the magnetic field gradient were measured to be up to Bz=0.495 T and up to Gz=2.2 T/m for the PF and the SF, respectively (see Fig. 1 LEFT). The MRI gradient strengths were determined to be Gx=0.84 mT/A/m, Gy=0.9 mT/A/m, Gz=0.87 mT/A/m (Imax=300 A) and the MRI RF coil was tuned to the Larmor frequency of protons at 0.495 T. The DF resonant circuits were tuned to excitation frequencies of around 25 kHz. With expected maximal DF amplitude of 20 mT and the strongest SF gradient strength a field of view of 18×36×36 mm3 could be reached.
机译:这种混合式MPI-MRI成像系统(见图1右)目前正在Bruker BioSpin MRI GmbH的研究实验室中组装,而单个组件已经过测试。 PF和SF的磁场强度和磁场梯度测得分别为 Bz = 0.495 T和G z = 2.2 T / m,分别(见图1左)。 MRI梯度强度确定为G x = 0.84 mT / A / m,G y = 0.9 mT / A / m,G z = 0.87 mT / A / m(I max = 300 A),MRI RF线圈被调谐到质子在0.495 T处的拉莫尔频率。 25 kHz。在预期最大DF振幅为20 mT和最强SF梯度强度的情况下,可以达到18×36×36 mm 3 的视场。