首页> 外文会议>International Workshop on Magnetic Particle Imaging >Receive coil optimization for an open magnetic particle imaging scanner
【24h】

Receive coil optimization for an open magnetic particle imaging scanner

机译:开放式磁粉成像扫描仪的接收线圈优化

获取原文

摘要

Helmholtz-like coil pairs with a radius of 15 cm have the highest sensitivity for the given geometry. Their homogeneity decreases with higher radius, which is both shown in Fig. 4. Arrays of circular coils show very good homogeneity, which further increases with increasing number of coils, but have comparably poor sensitivity in the FOV. D-shaped coils have both decreasing sensitivity and homogeneity with increasing radius. An increasing distance between a D-coil pair increases the homogeneity, but leads to decreasing sensitivity. Nested D-coils can show an improved homogeneity without a loss in sensitivity. Optimized circular and not nested D-shaped coil pairs with a radius of 15 cm were manufactured and the resonance frequency for both geometries was measured to be about 1.9 MHz for the tight and 2.6 MHz for the loose winding technique.
机译:对于给定的几何形状,半径为15 cm的类亥姆霍兹线圈对具有最高的灵敏度。它们的均匀性随着半径的增大而降低,这两者均如图4所示。圆形线圈阵列显示出非常好的均匀性,随着线圈数量的增加,均匀性进一步提高,但对FOV的敏感度却相对较差。 D形线圈随着半径的增加,灵敏度和均匀性均降低。 D线圈对之间距离的增加会增加均匀性,但会导致灵敏度降低。嵌套的D线圈可以显示出更高的均匀性,而不会降低灵敏度。制造出半径为15 cm的优化圆形和非嵌套D形线圈对,对于紧密配合,两种几何形状的谐振频率测得分别约为1.9 MHz和2.6 MHz。