Compensation method of frquency dependent attenuation for tissue harmonic imaging

机译:组织谐波成像的频率依赖性衰减补偿方法

获取原文

摘要

High definition of medical ultrasound images is strongly required. Recently, studies for ultrasound imaging using second harmonics have been advanced for this purpose. However, a living body has the property of frequency dependent attenuation (FDA). FDA is a problem that cannot be ignored when using the second harmonics for imaging. In this study, we propose an FDA compensation method for second harmonic pulse compression imaging in which ideal formed FM chirp echoes are received after having the distortion due to FDA by modulating the amplitude of a transmission signal.
机译:强烈需要高清晰度的医疗超声图像。 最近,为此目的,使用第二次谐波进行超声成像的研究。 然而,活体具有频率依赖性衰减(FDA)的性质。 FDA是在使用第二次谐波进行成像时不能忽略的问题。 在本研究中,我们提出了一种用于二次谐波脉冲压缩成像的FDA补偿方法,其中通过调制传输信号的幅度而具有失真之后接收理想形成的FM啁啾回波。