首页> 外文会议>IEEE International Ultrasonics Symposium >Concentration profiling using a novel acoustic backscatter system with single transducers pulsed at multiple frequencies
【24h】

Concentration profiling using a novel acoustic backscatter system with single transducers pulsed at multiple frequencies

机译:使用具有单个换能器的新型声反向散射系统浓缩分析,单个换能器以多个频率脉冲

获取原文

摘要

Acoustic backscatter systems offer a flexible technique to measure dispersion concentration and particle size, through inversion of the return echo voltage response of pulse-echo signals in the MHz range. This study explores the use of the ultrasound array research platform to measure the suspended particle attenuation constant using acoustic backscatter from glass species, and invert the voltage response to produce concentration profiles. The measurement principle is described and experimentally determined values for the particle attenuation constant are compared to theoretical values from literature. Transducer constants for each probe are determined allowing concentration profiles to be inverted from the voltage profiles, and the limits for determining the attenuation constant are explored.
机译:声反向散射系统提供一种灵活的技术来测量分散浓度和粒度,通过MHz范围内的脉冲回波信号的返回回波电压响应的反转来测量分散浓度和粒度。本研究探讨了超声阵列研究平台的使用,使用玻璃物质的声反向散射测量悬浮粒子衰减常数,并反转电压响应以产生浓度型材。将测量原理描述,并将颗粒衰减常数的实验确定的值与文献的理论值进行比较。每个探针的换能器常数被确定允许从电压轮廓倒置的浓度分布,并且探讨了确定衰减常数的限制。