High intensity focused ultrasound (HIFU) is a minimally invasive treatment modality for cancerous tumors. The monitoring of HIFU treatment is extremely important. Localized motion imaging (LMI) is a HIFU monitoring method basing on Harmonic motion method (HMI). It estimates the coagulation length by detecting the change of tissue stiffness. However, the previous LMI is not robust to noise, thus two techniques are employed to improve its noise robustness in this study. First, dynamic cross correlation window (DCCW) is used to obtain vibration amplitude with noise suppression. Second, vibration-frequency band pass filter (VBPF) is employed to calculate the amount of vibration with expected frequency. The qualitative evaluation indicates that the two techniques are able to suppress the affection of noise on the vibration amplitude and vibration amount, respectively. Furthermore, the quantitative comparison demonstrates that the improved LMI is able to give higher accuracy of coagulation length estimation than the previous LMI, and it reduces the root mean square error (RMSE) from 2.52 to 1.84 mm. In this article, an improved noise robust LMI is proposed for monitoring HIFU treatment.
机译:高强度聚焦超声(HIFU)是癌症肿瘤的微创治疗方式。监测HIFU治疗非常重要。局部运动成像(LMI)是基于谐波运动方法(HMI)的HIFU监测方法。它通过检测组织刚度的变化来估计凝血长度。然而,之前的LMI对噪声并不稳健,因此采用两种技术来改善本研究中的噪声鲁棒性。首先,使用动态互相关窗口(DCCW)来获得具有噪声抑制的振动幅度。其次,采用振动频带通过滤波器(VBPF)来计算具有预期频率的振动量。定性评估表明,这两种技术能够分别抑制噪声对振动幅度和振动量的影响。此外,定量比较表明,改进的LMI能够提供比前一个LMI更高的凝固长度估计精度,并且将根均方误差(RMSE)从2.52降至1.84mm。在本文中,提出了一种改进的噪声稳健LMI来监测HIFU治疗。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号