Dental implants are commonly used for oral rehabilitation. However, risks of failure still exist and are difficult to anticipate. This ex vivo study aims at investigating the potentiality of quantitative ultrasound (QUS) to assess the amount of bone in contact with titanium prototype cylindrical implants. Four groups of ten rabbit femurs each are considered, corresponding to specimen with different amounts of bone in contact with the implant. The 10 MHz ultrasonic response of the implant is processed to derive a quantitative indicator I, based on the temporal variation of the signal amplitude. ANOVA (p<10~(-5)) tests revealed a significant correlation between I and the amount of bone in contact with the cylinders. An empirical model considering the propagation of lateral waves allows to understand the physical origin of the echoes. QUS technique may be used to investigate the amount of bone in contact with cylindrically shaped implants.
机译:牙科植入物通常用于口腔康复。然而,失败风险仍然存在,并且很难预测。此外,前体内研究旨在调查定量超声(QUS)的潜力,以评估与钛原型圆柱形植入物接触的骨量。每组十组十个兔股骨被认为,对应于与植入物接触不同量的骨的样品。基于信号幅度的时间变化,处理植入物的10MHz超声波响应以导出定量指示器I。 ANOVA(P <10〜(-5))测试显示I与与气缸接触的骨骼之间的显着相关性。考虑横向波的传播的经验模型允许了解回波的物理来源。 QUS技术可用于研究与圆柱形植入物接触的骨的量。