Detection of the spatial accuracy of an O-Arm in the region of surgical interest

机译:检测外科兴趣区域的O形臂的空间精度

获取原文

摘要

Medical imaging is an essential component of a wide range of surgical procedures. For image guided surgical (IGS) procedures, medical images are the main source of information. The IGS procedures rely largely on obtained image data, so the data needs to provide differentiation between normal and abnormal tissues, especially when other surgical guidance devices are used in the procedures. The image data also needs to provide accurate spatial representation of the patient. This research has concentrated on the concept of accuracy assessment of IGS devices to meet the needs of quality assurance in the hospital environment. For this purpose, two precision engineered accuracy assessment phantoms have been developed as advanced materials and methods for the community. The phantoms were designed to mimic the volume of a human head as the common region of surgical interest (ROSI). This paper introduces the utilization of the phantoms in spatial accuracy assessment of a commercial surgical 3D CT scanner, the O-Arm. The study presents methods and results of image quality detection of possible geometrical distortions in the region of surgical interest. The results show that in the pre-determined ROSI there are clear image distortion and artefacts using too high imaging parameters when scanning the objects. On the other hand, when using optimal parameters, the O-Arm causes minimal error in IGS accuracy. The detected spatial inaccuracy of the O-Arm with used parameters was in the range of less than 1.00 mm.
机译:医学成像是各种外科手术的重要组成部分。对于图像引导的外科手术(IGS)程序,医学图像是信息的主要来源。 IGS程序在很大程度上依赖于获得的图像数据,因此数据需要在正常和异常组织之间提供分化,尤其是当在程序中使用其他外科手术设备时。图像数据还需要提供患者的准确空间表示。该研究集中了解IGS器件的准确性评估概念,以满足医院环境质量保证的需求。为此目的,两种精确的工程精度评估幻影被开发为社区的先进材料和方法。设计用于模拟人头的体积作为手术兴趣(ROSI)的常见区域。本文介绍了幻影在商业外科3D CT扫描仪的空间精度评估中的利用。该研究介绍了在手术兴趣区域中可能的几何扭曲的图像质量检测的方法和结果。结果表明,在预定的ROSI中,在扫描对象时,使用过高的成像参数存在明显的图像失真和伪影。另一方面,在使用最佳参数时,O形臂在IGS精度下会导致最小的误差。具有二手参数的O形臂的检测到的空间不准确度在小于1.00mm的范围内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号