Microphone placement evaluation for acoustic detection of coronary artery disease

机译:麦克风放置评估用于冠状动脉疾病的声学检测

获取原文
获取外文期刊封面目录资料

摘要

Coronary artery disease (CAD) is the leading cause of death in the United States yet no screening tool exists. The acoustic approach detects CAD noninvasively with microphones placed at the chest. Identifying ideal microphone placement on the body may improve accuracy. Here we examine measurements from acoustic recordings taken on the chest at three positions: the base, middle and apex level of the heart. Eight CAD positive and 13 negative subjects were used. Approximate entropy (ApEn) and largest Lyapunov exponents (LLE) were used to analyze measurements. Using these features the apex recording produced 88%–54% sensitivity-specificity while the middle and base positions produced 63%–77% and 0%–92% sensitivity-specificity respectively. These results show LLE and ApEn to be sensitive to CAD related signals and suggest that the best location for analysis is above the apex of the heart.
机译:在美国,冠状动脉疾病(CAD)是主要的死亡原因,但尚无任何筛查工具。声学方法是通过放置在胸部的麦克风无创地检测CAD。确定理想的麦克风放置位置可以提高准确性。在这里,我们检查了从胸部三个位置进行的声音记录得出的测量值:心脏的基础,中层和顶点。使用了八名CAD阳性和13名阴性受试者。使用近似熵(ApEn)和最大Lyapunov指数(LLE)来分析测量结果。使用这些功能,先端录音产生了88%–54%的敏感性特异性,而中间位置和基础位置分别产生了63%–77%和0%–92%的敏感性特异性。这些结果表明,LLE和ApEn对CAD相关信号敏感,表明分析的最佳位置是在心脏的最高点。