A dedicated processor for Monte Carlo computation in radiotherapy

机译:放射治疗中用于蒙特卡洛计算的专用处理器

获取原文

摘要

A dedicated processor for Monte Carlo computation in radiotherapy treatment planning on FPGA based hardware is being developed at INFN, Cagliari. When performing Monte Carlo simulations of the radiation dose delivered to the human body, the Compton interaction of a photon with an electron is simulated. A fast, pipelined, cost effective design for real time simulation of the Compton interaction had been implemented. The inputs to the system are the energy and the normalized direction vectors of the incoming photon. The energy and the direction vectors of the scattered photon and the scattered electron are calculated. The energy distribution by the scattered electron along its path in a voxel space is then calculated which can be used to construct maps of dose distribution in real time.
机译:Cagliari的INFN正在开发一种专用处理器,用于在基于FPGA的硬件上进行放射治疗治疗计划中的Monte Carlo计算。在对传递到人体的辐射剂量进行蒙特卡洛模拟时,将模拟光子与电子的康普顿相互作用。已经实现了一种快速,流水线,具有成本效益的设计,用于实时仿真Compton交互。系统的输入是入射光子的能量和归一化方向矢量。计算散射光子和散射电子的能量和方向矢量。然后,计算散射电子沿着其在体素空间中的路径的能量分布,该能量分布可用于实时构建剂量分布图。