首页> 外文会议>2010 IEEE Nuclear Science Symposium : Conference Record >Development of two-dimensional differential calibration method for a neutron dosimeter using a thermal neutron beam
【24h】

Development of two-dimensional differential calibration method for a neutron dosimeter using a thermal neutron beam

机译:用热中子束开发中子剂量计的二维差分校准方法

获取原文

摘要

We developed a new thermal neutron calibration method to be able to experimentally determine an energy response function of a neutron detector using a pulse parallel beam and the time-of-flight (TOF) technique. The calibration method was experimentally demonstrated for a 3He proportional counter and an electric personal dosimeter using a pulsed thermal neutron beam from the research reactor, JRR-3M. The responses of the detectors were successfully obtained as a function of the neutron energy. To solve the point at issue for the calibration method, we also propose an idea of an improved calibration method using a pulsed narrow beam, the TOF technique and a beam scanning technique. The improved calibration method does not require detailed information of the detector structure.
机译:我们开发了一种新的热中子校准方法,能够使用脉冲平行束和飞行时间(TOF)技术通过实验确定中子探测器的能量响应函数。使用来自研究堆JRR-3M的脉冲热中子束,通过实验证明了 3 He比例计数器和电子个人剂量计的校准方法。根据中子能量,成功获得了探测器的响应。为了解决校准方法存在的问题,我们还提出了一种使用脉冲窄光束,TOF技术和光束扫描技术的改进校准方法的想法。改进的校准方法不需要检测器结构的详细信息。