Validation of a new deterministic transport code for SPECT simulation

机译:验证用于SPECT模拟的新的确定性传输代码

获取原文

摘要

The simulation of single photon emission computed tomography (SPECT) has traditionally been done using Monte Carlo methods. However, the hybrid deterministic transport code TITAN is being benchmarked for the simulation of SPECT. The TITAN code is referred to as “hybrid” because it uses a discrete ordinates method in the phantom and a simplified ray-tracing algorithm in the air outside of the phantom. The TITAN code has been compared with the results of the SIMIND Monte Carlo code and experimental data of a myocardial perfusion phantom. In comparison with SIMIND, TITAN projection images were found to be in good visual agreement. However, maximum relative differences ranging from 13% to 26% were found when the projection images were compared numerically. These are believed to be the result of different cross-section information used in the codes and differences in how the codes represent collimation. A computation time comparison showed the advantage of the TITAN code over SIMIND when an increasing number of projection images are generated. The comparison with experimental data also resulted in excellent visual agreement with numerical differences of 21% to 30%, which are at least partially due to limitations of the experiment.
机译:传统上,使用蒙特卡洛方法完成了单光子发射计算机断层扫描(SPECT)的仿真。但是,混合确定性传输代码TITAN正在作为SPECT模拟的基准。 TITAN代码被称为“混合”代码,因为它在体模中使用离散坐标方法,并在体模外部的空气中使用简化的光线跟踪算法。 TITAN代码已与SIMIND Monte Carlo代码的结果以及心肌灌注模型的实验数据进行了比较。与SIMIND相比,TITAN投影图像具有良好的视觉一致性。但是,当对投影图像进行数值比较时,发现最大相对差异在13%到26%之间。据信,这是由于编码中使用的横截面信息不同以及编码表示准直的方式不同而导致的。计算时间比较显示,当生成越来越多的投影图像时,T​​ITAN代码优于SIMIND。与实验数据的比较还产生了极好的视觉一致性,其数值差异为21%至30%,这至少部分是由于实验的局限性所致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号