Multi-channel magnetic induction tomography measurement system

机译:多通道磁感应层析成像测量系统

获取原文

摘要

Magnetic induction tomography(MIT) is a noninvasive and contactless medical imaging technique. The 16 channels of MIT system is showed in the paper, which image for the low conductivity object with working at 1MHz. In the study, we use the water tanker filled with salt solution and the different conductivity agar blocks to simulate the human brain with edema or hematoma. The each sensor is composed of the excitation coil, the two detection coils, a preamplifier board and a power amplifier board, which were fixed by four plastic columns. This paper researches the sensitivity and stability of single sensor unit, and the phase sensitivity of the system is 0.29°/S·m−1 and the phase drift is in the range of −0.02° to +0.06° over 30 minutes. Finally, the 2-dimensional images of conductivity distribution are obtained using the inverse distance weighted spatial interpolation algorithm of and the direct projection way.
机译:磁感应断层扫描(MIT)是一种非侵入性和非接触式医学成像技术。本文显示了MIT系统的16个通道,这些通道是工作在1MHz的低电导率对象的图像。在研究中,我们使用装满盐溶液的水槽和不同电导率的琼脂块来模拟人脑水肿或血肿。每个传感器由励磁线圈,两个检测线圈,一个前置放大器板和一个功率放大器板组成,它们由四个塑料柱固定。本文研究了单个传感器单元的灵敏度和稳定性,系统的相位灵敏度为0.29°/ S·m -1 ,相位漂移在-0.02°至+0.06范围内°30分钟以上。最后,采用直接投影方式的反距离加权空间插值算法获得了电导率分布的二维图像。