【24h】

CZT/NaI hybrid hand-held performance evaluation

机译:CZT / NaI混合手持性能评估

获取原文

摘要

New hybrid spectroscopic systems directly combine spectra from detectors with very different energy resolutions, accommodating standard analyses of the output hybrid spectrum. Hand-held hybrid systems consisting of a 0.5 cm3 cadmium zinc telluride detector combined with either a 1 cm3 NaI(Tl) scintillator mounted on a photomultiplier tube or a 1 cm3 CsI(Tl) mounted on a silicon photomultiplier were evaluated for performance by source measurements for a variety of acquisition times. Repeated 300 second background measurements were analyzed and found a small increase in false alarm rate for the hybrid combinations compared to the individual component detectors. Repeated measurements with a 100 μR/h 137Cs source and acquisition times of 5, 10, 30, 100, and 300 seconds were analyzed, with the results showing that hybridization significantly enhances peak detectability relative to the individual constituent detectors, especially at shorter times. At 5 and 10 seconds, the probability of detection was more than twice that of the individual components. The hybrid approach enables the consideration of a greater variety of measurement system solutions in terms of cost and performance.
机译:新的混合光谱系统直接将光谱与具有非常不同的能量分辨率的探测器相结合,适应输出混合谱的标准分析。由0.5cm 3 镉碲化镉探测器组成的手持式混合系统与安装在光电倍增管或1cm上的1cm 3 nai(tl)闪烁体组合通过源测量评估安装在硅光电倍增管上的3 CSI(TL),通过源测量进行各种采集时间。与各个组件探测器相比,分析了重复300秒的背景测量并发现混合组合的误报率的小幅增加。分析了100μR/ h 137 cs源和5,10,30,100和300秒的测量的重复测量,结果表明杂交显着提高了相对于个体的峰值可检测性构成探测器,尤其是较短的时间。在5和10秒,检测概率超过各个组分的两倍多。混合方法能够在成本和性能方面考虑更多各种测量系统解决方案。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号