Experimental Study of the Pressure Reduction on AAA Wall by the Insertion of a Stent-graft

机译:支架植入术对AAA壁减压的实验研究

获取原文

摘要

A set of experiment apparatus simulating human heart and artery were set up, including an elastic Abdominal Aortic Aneurysm (AAA) model made of rubber. The healing influence on an AAA resulted from the stent-grafts with different diameters was investigated. The experimental results show that the pressure on AAA wall is reduced after the insertion of a stent-graft. The porosity of a stent-graft is not proportionally related to the pressure reduction. Additionally, the reduction of pressure is affected by the stent-graft diameter: the stent-graft in the diameter of being 10~15 percent larger than the artery diameter can reduce the wall pressure in the most effective way- The results are expected to have important application in the clinical.
机译:建立了一套模拟人的心脏和动脉的实验设备,包括一个由橡胶制成的弹性腹主动脉瘤(AAA)模型。研究了不同直径的支架移植物对AAA的愈合影响。实验结果表明,在植入支架后,AAA壁上的压力会降低。支架移植物的孔隙率与压力降低不成比例。此外,压力的降低还受到支架移植物直径的影响:直径大于动脉直径10%到15%的支架移植物可以最有效地降低壁压-预期结果是在临床上的重要应用。