首页> 外文会议>Second International Conference on Coordinated and Multiple Views in Exploratory Visualization, 2004. Proceedings, 2004 >Electronic image archiving in the catheterization laboratory:multicenter comparison of enhanced optical storage with 10242digital and 35 mm cine-film
【24h】

Electronic image archiving in the catheterization laboratory:multicenter comparison of enhanced optical storage with 10242digital and 35 mm cine-film

机译:导管实验室中的电子图像存档:具有1024 2 数字和35毫米电影胶片的增强光学存储的多中心比较

获取原文

摘要

The authors developed a bi-directional 1249/625 lines converter,able to store and to replay high resolution (1249 lines) video images,from Component Record Video (CRV) Optical Laser (Analogue) Videodisks.Image quality was evaluated four ways: (1) The authors evaluatedQuantitative Coronary Angiography (QCA) on standard coronary testphantoms using enhanced CRV storage images digitally processed and alsocompared 78 arterial measurements ranging 0.65 to 4.85 mm both front thedigital hard disc (D) and the CRV optical disc (CRV), using the same QCAanalytical package. No systematic over or underestimation occurred, themean variability was inferior to 0.1 mm and results were similar to theresults obtained in the literature from digitally stored images. (2)Outside experts compared 593 static images stored on CRV discs to thesame images stored in 10242 digital format blinded to sourceof image. The four experts found no visual or medical difference in 98%of evaluated images and minor differences in the remainder. (3) 90representatives from 63 institutions compared images from both a digitalhard drive and the enhanced CRV optical (analogue) storage. During thetest they were blinded as to the source of the images. Therepresentatives found no difference in image resolution, qualitydiagnostic accuracy, and medical relevance. (4) Two of the experts alsocompared the enhanced CRV optically stored image to the image obtainedon simultaneously recorded 35 mm cine film, and found the enhanced CRVstored image to be superior or similar, but never inferior to the filmimage. It is concluded that High-Resolution CRV storage represents acost effective solution for excellent image quality equivalent todigitally stored images, permitting permanent electronic archivinginside the cath-lab, and allowing digital image processing and digitalDICOM network and CD-ROM based image communication
机译:作者开发了一个双向1249/625线转换器, 能够存储和重播高分辨率(1249行)视频图像, 来自组件记录视频(CRV)光学激光(模拟)视频电视程。 图像质量评估了四种方式:(1)作者评估 定量冠状动脉造影(QCA)标准冠状动脉试验 使用增强型CRV存储图像的幻像数字处理和此外 比较了78个动脉测量值0.65至4.85mm前面 数字硬盘(D)和CRV光盘(CRV),使用相同的QCA 分析包。没有发生系统的过度或低估 平均变异性低于0.1mm,结果类似于 从数字存储图像中的文献中获得的结果。 (2) 外部专家比较593静态图像,存储在CRV光盘上 存储在1024 2 数字格式中的相同图像被蒙蔽源 图像。这四名专家发现了98%的视觉或医学差异 评估的图像和剩余部分的微小差异。 (3)90 来自63个机构的代表比较了数字 硬盘驱动器和增强型CRV光学(模拟)存储。在此期间 测试它们被蒙蔽了图像的来源。这 代表发现图像分辨率没有区别,质量 诊断准确性和医学相关性。 (4)两个专家也 将增强的CRV光学存储图像与获得的图像进行了比较 同时记录了35毫米的电影,并发现增强的CRV 存储的图像是优越的或相似的,但从不差不多到电影 图像。得出结论,高分辨率CRV存储代表一个 具有优异的图像质量的成本有效解决方案 数字存储的图像,允许永久电子归档 在Cath-Lab内,并允许数字图像处理和数字 DICOM网络和基于CD-ROM的图像通信

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号