A digital signal processor for Doppler radar sensing of vital signs

机译:用于多普勒雷达生命体征的数字信号处理器

获取原文

摘要

A voltage waveform signal containing respiration and heartbeat signatures is low-pass filtered (0.7 Hz) for the respiration and band-pass filtered (1.0-3.0 Hz) for the heart signal. The autocorrelation function is used to calculate the rate per minute for each of these signals. To make the processor more robust, several signal processing techniques are applied. One of these techniques, a method commonly used on audio signals for formant removal, is called center clipping. Another technique is the use of a Hanning window. To reverse the distinct shape left by the Hanning window an "undo" window is applied. The processor is programmed in LabVIEW.
机译:包含呼吸和心跳信号的电压波形信号经过低通滤波(0.7 Hz)进行呼吸,并经过带通滤波(1.0-3.0 Hz)进行心脏信号通行。自相关函数用于计算这些信号中每一个的每分钟速率。为了使处理器更强大,应用了几种信号处理技术。这些技术中的一种,通常用于音频信号以去除共振峰的方法,称为中心削波。另一种技术是使用Hanning窗口。为了反转汉宁窗留下的独特形状,应用了“撤消”窗。该处理器是在LabVIEW中编程的。