首页> 外文会议>Instrumentation and Measurement Technology Conference, 2000. IMTC 2000. Proceedings of the 17th IEEE >Speckle velocimeter with a self-mixing laser diode: optimization of measurement system, reproducibility and accuracy of measurements
【24h】

Speckle velocimeter with a self-mixing laser diode: optimization of measurement system, reproducibility and accuracy of measurements

机译:带自混合激光二极管的散斑测速仪:测量系统的优化,测量的可重复性和准确性

获取原文

摘要

A self-mixing laser diode speckle velocimeter based on autocorrelation of speckle signals is analyzed in detail. In this velocimeter scheme, a semiconductor laser is used both as the source and the detector enabling a self detecting, self-aligning system with a single optical axis. Effects of different parameters (beam spot size, measurement time, etc) of the measurement system are analyzed and the system was optimized for the best accuracy and reproducibility of the measurements. The optimized system has a measurement error less than 1.5% within the range of calibration velocities and less than 3% out of this range.
机译:详细分析了基于散斑信号自相关的自混合激光二极管散斑速度仪。在这种测速仪方案中,半导体激光器既用作光源,又用作探测器,可实现具有单个光轴的自检测,自对准系统。分析了测量系统的不同参数(光束斑大小,测量时间等)的影响,并对系统进行了优化,以实现最佳的测量精度和可重复性。经过优化的系统在校准速度范围内的测量误差小于1.5%,在该范围之外的测量误差小于3%。