Abstract: The overall goal of our program is to develop a robust, high throughput, fully automated DNA sequencing instrument based on replaceable polymer solutions using a multicapillary array. Significant effort has already been devoted to column and polymer chemistry in order to obtain long read lengths per run in fast analysis time. In this paper we report on progress in instrument considerations and data processing software. A simple instrument design, based on no moving parts for continuous illumination of the capillaries and detection of the fluorescent light was used for this study. Our polymer solution replacement system with the permanent connection between the buffer/chamber manifold and capillary columns on the detector side is designed to prevent the trapping of air bubbles during matrix solution replacement. A special construction of a column-electrode couple on the injection side precludes air trapping during sample injection from small sample volumes. Our in-house software now features the significant reduction of the crosstalk signal from neighbor columns, which may be a potential problem in densely packed large capillary array sequencers. !20
机译:摘要:我们计划的总体目标是,基于使用多毛细管阵列的可替换聚合物解决方案,开发一款功能强大,高通量的全自动DNA测序仪。为了在快速的分析时间内每次运行获得较长的读取长度,已经在色谱柱和聚合物化学上投入了大量精力。在本文中,我们报告了仪器注意事项和数据处理软件方面的进展。本研究使用了一种基于无移动部件的简单仪器设计,以连续照明毛细管并检测荧光。我们的聚合物溶液替换系统在检测器侧的缓冲液/腔室歧管和毛细管柱之间具有永久连接,旨在防止基质溶液替换期间残留气泡。在进样侧采用特殊的柱电极耦合结构,可防止在进样过程中从小体积样品中捕获空气。我们的内部软件现在具有显着降低来自相邻色谱柱的串扰信号的功能,这在密集包装的大型毛细管阵列测序仪中可能是一个潜在问题。 !20

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号