【24h】

Arrhythmia sensing and detection by a software-based arrhythmia control system

机译:通过基于软件的心律不齐控制系统进行心律不齐感测和检测

获取原文

摘要

A microprocessor-based arrhythmia control system (ACS) has been developed to provide comprehensive automatic control of certain cardiac arrhythmias, including ventricular bradycardia, ventricular tachycardia (VT), and ventricular fibrillation (VF). A dual sensing system, three tachycardia detection algorithms, and two tachycardia confirmation algorithms are used to sense, detect, and confirm cardiac arrhythmias. These algorithms and the sensing system of the ACS have been tested with 27 VT/VF episodes of human intracardiac electrograms. The results demonstrated that the ACS can sense, detect, and confirm human cardiac arrhythmias appropriately.
机译:已开发出基于微处理器的心律不齐控制系统(ACS),以提供对某些心律不齐的全面自动控制,包括心室心动过缓,室性心动过速(VT)和心室纤颤(VF)。双传感系统,三种心动过速检测算法和两种心动过速确认算法用于感测,检测和确认心律不齐。这些算法和ACS的感应系统已通过人心内电描记图的27次VT / VF发作进行了测试。结果表明,ACS可以适当地感应,检测和确认人的心律不齐。