11 at the front of the sample with a coaxial-feed type cut-off waveguide was measu'/> Dielectric Measurement in Liquids via the Cut-off Circular Waveguide Reflection Method After Sil Calibration Using SOM Termination
首页> 外文会议>International Microwave and Radar Conference >Dielectric Measurement in Liquids via the Cut-off Circular Waveguide Reflection Method After Sil Calibration Using SOM Termination
【24h】

Dielectric Measurement in Liquids via the Cut-off Circular Waveguide Reflection Method After Sil Calibration Using SOM Termination

机译:使用SOM端接进行Sil标定后,通过截止圆波导反射法测量液体中的介电常数

获取原文

摘要

In this study, S11 at the front of the sample with a coaxial-feed type cut-off waveguide was measured in the frequency band of 0.50 to 3.0 GHz with liquids in the cut-off circular waveguide after S11 calibration using SOM (short, open and reference material) conditions. The permittivity of liquids was estimated from the above S11 based on 1. an inverse problem approach involving comparison with the S11 value calculated using electromagnetic analysis via the mode-matching method, and 2. an estimation formula involving comparison with termination associated with SOM conditions. The Debye dispersion equation for calculation of methanol’s complex permittivity was corrected from the results. The dielectric constants of various liquids were estimated using a formula for comparison with three reference materials (pure water, methanol and air) using methanol after dielectric-constant correction. The effectiveness of the proposed method was verified by comparing the above estimated values with otherwise obtained outcomes. Moreover, the S11 value of the jig was calibrated with short, open and another reference material (methanol) conditions after permittivity correction of methanol. The estimation results for the dielectric constant of pure water and similar after calibration were also compared with the above estimation values, and the reversibility of the proposed estimation method was verified. Favorable agreement of values estimated with each method was observed, indicating the validity of the proposed dielectric measurement procedure.
机译:在这项研究中,S 11 S后,在带有圆形馈电型截止波导的样品的前端,在0.50至3.0 GHz的频带中,在截止环形波导中有液体,测量了样品前部的 11 使用SOM(短材料,开放材料和参考材料)条件进行校准。液体的介电常数由上述S估算 11 基于1.基于反问题的方法,其中涉及与S的比较 11 使用通过模式匹配方法的电磁分析计算得出的值,以及2.与与SOM条件相关的终止进行比较的估计公式。从结果中校正了用于计算甲醇复介电常数的德拜弥散方程。介电常数校正后,使用甲醇与三种参比物质(纯水,甲醇和空气)进行比较,使用公式估算各种液体的介电常数。通过将上述估计值与以其他方式获得的结果进行比较,验证了所提方法的有效性。而且,S 11 在甲醇的介电常数校正后,用短,开和其他参考材料(甲醇)条件校准夹具的值。还将校准后纯水和类似物的介电常数的估算结果与上述估算值进行了比较,并验证了所提出估算方法的可逆性。观察到每种方法估计的值具有良好的一致性,表明所提出的介电测量程序的有效性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号