Purpose: Low levels of ghosting are found during quality control of digital mammography (DM). In digital breast tomosynthesis (DBT) mode, the x-ray detector is not cleared during the series of exposures potentially leading to ghosting issues. Methods: The ghosting image factor (GIF) was calculated for systems from three manufacturers as follows. A 45 mm thick polymethyl methacrylate (PMMA) block covering half the detector was imaged, followed by the PMMA block plus O.I mm thick aluminium sheet centred on the block, this time positioned to fully cover the detector. This was undertaken for (1) DM and (2) between separate DBT scans. Inter-projection ghosting (3) was measured by moving the PMMA block from covering half the detector to covering the whole detector after the first projection. In method (4), a 3 mm thick aluminium plate with a 0.1 mm thick aluminium sheet on top was placed on the compression paddle at 100 mm above the breast covering half of the detector. Results: The GIF measured via test (1)&(2) was lower than the 0.3 tolerance set in the European guidelines for DM, however higher values were found for one system in DBT mode. Test 3, inter-projection GIF values were between 0.08 and 0.89, mostly above 0.3. Test 4 was a practical method and gave larger GIF values of between 1.07 and 8.01. Conclusions Ghosting factors measured between projections are considerably higher than the GIF measured between separate DBT scans. We propose a simple method using an aluminium plate to estimate the GIF.
机译:目的:在数字乳腺摄影(DM)的质量控制过程中发现少量的重影。在数字乳房断层合成(DBT)模式下,在一系列曝光期间未清除X射线检测器,这可能会导致重影问题。方法:如下计算了来自三个制造商的系统的重影图像因子(GIF)。对覆盖一半检测器的45毫米厚的聚甲基丙烯酸甲酯(PMMA)块成像,然后是PMMA块加上以块为中心的0.1毫米厚的铝片,这次定位为完全覆盖检测器。这是针对(1)DM和(2)在独立DBT扫描之间进行的。投影间重影(3)通过在第一次投影后将PMMA块从覆盖探测器的一半移动到覆盖整个探测器的方式进行测量。在方法(4)中,将一块3毫米厚的铝板和一个顶部为0.1毫米厚的铝片放在压迫板上,在覆盖乳房一半的探测器上方100毫米处。结果:通过测试(1)&(2)测量的GIF低于欧洲DM准则中设置的0.3公差,但是在DBT模式下发现一个系统的GIF值较高。测试3,投影间的GIF值在0.08到0.89之间,大部分在0.3以上。测试4是一种实用的方法,其GIF值在1.07至8.01之间。结论在投影之间测得的重影因子大大高于在单独的DBT扫描之间测得的GIF。我们提出了一种使用铝板估算GIF的简单方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号