Clinical Validation of the Saadat NIBP Module According to the BS EN ISO 81060-2 Protocol

机译:根据BS EN ISO 81060-2协议对Saadat NIBP模块进行临床验证

获取原文

摘要

This study was conducted according to the British Standard EN ISO 81060-2 guidelines. The aim was to validate the Saadat non-invasive blood pressure (NIBP) module accuracy and reliability against manual auscultatory general practitioners (GP). readings. A total of 298 measurements and comparison procedures were performed on 95 adults without heart disease under the supervision of two GPs. In order to represent the relationships between the test device and the reference method, the Bland-Altman graphical plotting method was used. The mean differences and standard deviations (mean ± SD differences) between the readings of the Saadat NIBP module and determination of GPs as the reference method for systolic and diastolic were exceptionally close with: -2.22 ± 6.51 mmHg and -3.31 ± 6.27 mmHg respectively. The Saadat NIBP module fulfilled the BS EN ISO 81060-2 requirements, which states that the mean ± SD of lower than 5 ± 8 mmHg can be recommended for clinical use.
机译:这项研究根据英国标准EN ISO 81060-2指南进行。目的是验证萨迪茨无侵入性血压(NIBP)模块的准确性和可靠性对手动听诊的通用从业者(GP)。读物。在两个GPS的监督下,总共对95名成人进行了95例成人进行了298次测量和比较程序。为了代表测试装置与参考方法之间的关系,使用了平坦的altman图形绘制方法。 Saadat NIBP模块读数与作为收缩型和舒张的参考方法的GPS读数之间的平均差异和标准偏差(平均值±SD差异)特别接近:-2.22±6.51mmHg和-3.31±6.27mmHg。 Saadat NIBP模块满足了BS EN ISO 81060-2的要求,这些要求将推荐用于临床使用的平均值±8 mmHg。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号