首页> 外文会议>Society for Biomaterials annual meeting and exposition >Degradable Devices for Localized Delivery of Immunosuppressants to Prevent Transplant Rejection
【24h】

Degradable Devices for Localized Delivery of Immunosuppressants to Prevent Transplant Rejection

机译:用于局部递送免疫抑制剂以防止移植排斥的可降解装置

获取原文

摘要

TyrPC devices loaded with tacrolimus were successfully fabricated, sterilized, characterized, and implanted into animal models. Target drug concentrations were reached systemically for one week in vivo while localized tissue tacrolimus levels remained measureable at three weeks post-implantation. Together, these results indicate that the devices provide systemic immunosuppression initially and localized long-term. Future work will adjust device formulation and implant number to extend the therapy duration and study the extent of localized immunosuppression achieved.
机译:装有他克莫司的TyrPC装置已成功制造,消毒,定性并植入动物模型。在体内全身达到目标药物浓度一周,而在植入后三周仍可测量局部组织他克莫司水平。在一起,这些结果表明该设备最初提供系统的免疫抑制和长期定位。未来的工作将调整设备的配方和植入物的数量,以延长治疗时间并研究局部免疫抑制的程度。