This paper describes an R-R Interval (RRI) outlier exclusion method that can exclude RRI outliers including misdetected R waves caused by artifacts, which are frequently observed when using wearable electrocardiogram (ECG) devices. The method distinguishes targeted misdetected R waves based on statistical ECG values that can reflect the occurrence of artifacts, and annotates all detected R waves with those results for evaluating RRI measurement status. Experimental results confirmed that the proposed method is effective to improve the accuracy of both time and frequency domain measures of heart rate variability than the conventional one.
机译:本文介绍了一种R-R间隔(RRI)离群值排除方法,该方法可以排除RRI离群值,其中包括由伪影引起的误检测到的R波,在使用可穿戴心电图(ECG)设备时经常会观察到这种异常。该方法基于统计的ECG值来区分目标误检测的R波,该统计ECG值可反映伪像的出现,并用那些结果注释所有检测到的R波,以评估RRI测量状态。实验结果证实,与传统方法相比,该方法可以有效地提高心率变异性的时域和频域测量精度。