A Phantom Study of New Bias Field Correction Method Combining N3 and KHM for MRI Imaging

机译:结合N3和KHM进行MRI成像的新型偏场校正方法的幻象研究。

获取原文
获取外文期刊封面目录资料

摘要

Signal inhomogeneity in MRI can influence significantly automatic data processing like segmentation, etc. or even affect the diagnostic procedure. In this work, a new method of intensity nonuniformity correction is presented. Our idea was to replace FCM clustering by k-harmonic means in the method that uses a standard N3 correction procedure. The algorithm was tested with MRI dataset acquired from a phantom object using a breast MRI coil to simulate real conditions during the study. Results were compared with five other methods using two indexes-integral uniformity and standard deviation of the signal inside the object. For the presented and improved method, the lowest integral uniformity and the reasonable low signal deviation were obtained.
机译:MRI中的信号不均匀会严重影响自动数据处理(例如分段等),甚至会影响诊断过程。在这项工作中,提出了一种强度不均匀校正的新方法。我们的想法是,在使用标准N3校正程序的方法中,以k谐和方式代替FCM聚类。该算法通过使用MRI乳房成像线圈从幻像对象获取的MRI数据集进行了测试,以模拟研究过程中的实际情况。使用两个指标(对象的内部信号的整体均匀性和标准偏差),将结果与其他五种方法进行了比较。对于提出和改进的方法,获得了最低的积分均匀性和合理的低信号偏差。