Diabetes Mellitus is one of the notable health concerns around the globe. To keep diabetes in check a patient needs to track his/her blood glucose level regularly. In the traditional invasive system, a patient needs to prick his body part to collect the blood sample, which deems to be painful for the patient. Non-invasive blood glucose measurement can diminish the pain. In this paper, we proposed a non-invasive blood glucose determination method where a NIR LED (940nm) and a photo-detector is used to determine the blood glucose concentration. We used the diffused reflectance method to determine the blood glucose concentration in our paper. After implementing the device, we compared the data between our device and established invasive blood glucometer available in the market. Clarke grid error analysis shows, that most of the measured data by the device are within Region-A.
机译:糖尿病是全球著名的健康问题之一。为了控制糖尿病,患者需要定期追踪他/她的血糖水平。在传统的侵入式系统中,患者需要戳刺自己的身体部位以收集血液样本,这对患者来说很痛苦。无创血糖测量可以减轻疼痛。在本文中,我们提出了一种无创血糖测定方法,其中使用NIR LED(940nm)和光电检测器确定血糖浓度。我们使用了漫反射法来确定我们论文中的血糖浓度。实施该设备后,我们将设备与市场上已有的侵入式血糖仪之间的数据进行了比较。 Clarke网格误差分析表明,该设备的大多数测量数据都在Region-A区域内。