Hypochlorous Acid Sensor using Boron-Doped Diamond Electrode in Physiological pH Solution

机译:在生理pH溶液中使用掺硼金刚石电极的次氯酸传感器

获取原文

摘要

Electrochemical behavior of hypochlorous acid (HClO) have been studied at boron-doped diamond (BDD) electrodes in physiological pH solution. The electrochemical measurements were performed in one-compartment electrochemical cell using cyclic voltammetry technique. A solution of 0.5 M phosphate buffer solution has been selected as the supporting electrolyte. Cyclic voltammograms of NaClO standard solutions showed the best response at optimum pH of 6.6 at the potential reduction of +0.1 V. Linear calibration curve ( R2=0.9953) can be achieved at a NaClO concentration range from 1.68 to 5.86 mM with an estimated limit of detection (LOD) of 0.46 mM.
机译:次氯酸(HClO)的电化学行为已在生理pH溶液中的掺硼金刚石(BDD)电极上进行了研究。使用循环伏安法在单室电化学池中进行电化学测量。选择了0.5 M磷酸盐缓冲溶液作为支持电解质。 NaClO标准溶液的循环伏安图显示了在最佳pH值为6.6且电位降低+0.1 V时的最佳响应。线性校准曲线(R 2 在1.68至5.86 mM的NaClO浓度范围内可达到0.993),估计检出限(LOD)为0.46 mM。