Evaluation of an RF wearable device for non-invasive real-time hydration monitoring

机译:用于非侵入式实时水化监测的射频可穿戴设备的评估

获取原文

摘要

In this paper, we validate a novel wearable device for real time non-invasive hydration monitoring. An experiment was carried out on three sets of 12 subjects under various exercise intensities to identify how the received signal strength indicator (RSSI) captured with the device relates to the change in the body hydration, body water loss percentage (BWL%). A linear regression correlation R of 0.77 and a third degree polynomials correlation (R=0.83) between ΔRSSI and BWL% is established based on the observation of the experiment, which implies that the hydration status can be quantized as body water loss by measuring ΔRSSI. Therefore, the technique and device is verified to be a potentially solid solution for wearable non-invasive real-time hydration monitoring device.
机译:在本文中,我们验证了一种新颖的可穿戴设备,用于实时非侵入式水合作用监测。在不同运动强度下,对三组12名受试者进行了一项实验,以确定使用该设备捕获的接收信号强度指示器(RSSI)与体内水分变化,人体失水百分比(BWL \%)的关系。根据实验观察,建立了线性回归相关系数R为0.77和三阶多项式相关系数(R = 0.83),ΔRSSI与BWL \%之间存在关系,这表明可以通过测量ΔRSSI将水合作用状态量化为人体失水量。 。因此,该技术和设备被证实是可穿戴非侵入式实时水化监测设备的潜在固体解决方案。