Systolic time interval estimation at the sternum using continuous wave radar with body-contact antennas

机译:使用带有人体接触天线的连续波雷达估算胸骨的收缩时间间隔

获取原文

摘要

Cardiovascular vital signs are measured using continuous wave (CW) radar at 2.45GHz with a body-contact antenna. Using a previously collected database of thirty second recordings at paced breathing conditions, CW radar signals were measured together with heart sounds, electrocardiogram (ECG), respiration and impedance cardiogram (ICG) were as reference signals. Using arbitrarily chosen features from the radar signal, the systolic time intervals (STI)s from radar are compared with those from the ICG. The correlation coefficients between the STIs of radar and ICG were 0.72, 0.66 and 0.81 for the pre-ejection period, left ventricular ejection time and electromechanical systole respectively. The p-value was below 0.0001 for all coefficients, indicating a significant correlation. The results indicate that the radar signals capture cardiomechanical signals, which have great potential to be used for STI estimation in ambulatory conditions.
机译:使用带有身体接触天线的2.45GHz连续波(CW)雷达测量心血管生命体征。使用先前收集的在有节奏的呼吸条件下三十秒记录的数据库,将CW雷达信号与心音一起测量,心电图(ECG),呼吸和阻抗心电图(ICG)作为参考信号。使用从雷达信号中任意选择的特征,将雷达的脉动时间间隔(STI)与ICG的脉动时间间隔进行比较。在射血前期,左心室射血时间和机电收缩期,雷达与ICG的STI的相关系数分别为0.72、0.66和0.81。所有系数的p值均低于0.0001,表明存在显着相关性。结果表明,雷达信号捕获了心脏机械信号,具有很大的潜力可用于非卧床条件下的STI估计。