Evaluation of RIP sensor calibration stability for daily estimation of lung volume

机译:评估RIP传感器校准稳定性以每日估算肺活量

获取原文

摘要

This work supports the development of a non-invasive wearable system for the tracking of cigarette smoking in the natural environment. Accurate estimation of lung volume, as well as preservation of calibration over the period of wear, are required for such a system. In this paper, the preservation of calibration of a Respiratory Inductive Plethysmograph (RIP) in a 24-hour period was evaluated. Attachment options, such as an RIP belt held in place by tension (belt), RIP belt held in place by adhesive (belt+adhesive), and RIP belt sewn into a T-shirt (belt+shirt) were studied and compared using data collected from three subjects performing usual daily activities. An 800 ml bag was used as the gold standard for evaluation. The average volume obtained for the belt, belt+adhesive, and belt+shirt options were 0.89+0.28 L, 1.02+0.27 L, and 0.79+0.05 L, respectively. Additionally, comparing these against a reference volume of 800 ml, the average RMSE was 0.24+0.05 L, 0.24+0.23 L, and 0.03+0.03 L, respectively. Results show that the RIP belt sewn into a T-shirt provides a better preservation of the calibration throughout the 24-hour period.
机译:这项工作支持开发用于在自然环境中跟踪吸烟的非侵入性可穿戴系统。对于这样的系统,需要准确估计肺部容积以及在整个佩戴过程中保持校准。在本文中,评估了24小时内呼吸式呼吸描记器(RIP)的校准保存。研究了附加选项,例如通过张力(皮带)固定在位的RIP皮带,通过粘合剂(皮带+粘合剂)固定在位的RIP皮带和缝入T恤(皮带+衬衫)的RIP皮带,并使用数据进行了比较从进行日常日常活动的三名受试者中收集。使用800毫升袋作为评估的金标准。皮带,皮带+粘合剂和皮带+衬衫选项的平均体积分别为0.89 + 0.28 L,1.02 + 0.27 L和0.79 + 0.05L。另外,将它们与800 ml的参考体积进行比较,平均RMSE分别为0.24 + 0.05 L,0.24 + 0.23 L和0.03 + 0.03L。结果表明,缝在T恤衫中的RIP皮带可在整个24小时内更好地保存校准数据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号