The evaluation of correction algorithms of intensity nonuniformity in breast MRI images: a phantom study

机译:乳腺MRI图像强度不均匀校正算法的评估:一项幻像研究

获取原文

摘要

The aim of this work was to test the most popular and essential algorithms of the intensity nonuniformity correction of the breast MRI imaging. In this type of MRI imaging, especially in the proximity of the coil, the signal is strong but also can produce some inhomogencitics. Evaluated methods of signal correction were: N3, N3FCM, N4, Nonparametric, and SPM. For testing purposes, a uniform phantom object was used to obtain test images using breast imaging MRI coil. To quantify the results, two measures were used: integral uniformity and standard deviation. For each algorithm minimum, average and maximum values of both evaluation factors have been calculated using the binary mask created for the phantom. In the result, two methods obtained the lowest values in these measures: N3FCM and N4, however, for the second method visually phantom was the most uniform after correction.
机译:这项工作的目的是测试乳房MRI成像强度不均匀校正的最流行和最基本的算法。在这种类型的MRI成像中,尤其是在线圈附近,信号很强,但也会产生一些非均质性。评估信号校正的方法是:N3,N3FCM,N4,非参数和SPM。出于测试目的,使用统一的幻影物体使用乳房成像MRI线圈获得测试图像。为了量化结果,使用了两种方法:积分均匀性和标准偏差。对于每种算法,已经使用为模型创建的二进制掩码计算了两个评估因子的最小值,平均值和最大值。结果,两种方法在这些度量中获得了最低的值:N3FCM和N4,但是,对于第二种方法,在校正后,视觉幻像是最均匀的。