首页> 外文会议>International conference on robotic, vision, signal processing and power applications >Deflection Analysis of a Robotic Bed on the Applied Loads and Its Postures for a Heavy-Ion Therapeutic System
【24h】

Deflection Analysis of a Robotic Bed on the Applied Loads and Its Postures for a Heavy-Ion Therapeutic System

机译:机器人床在重离子治疗系统上施加的载荷及其姿势的挠度分析

获取原文

摘要

In this paper, we present two preliminary experiments, under laboratory environment, in order to determine whether or not our robotic bed needs a deflection correction function due to patient's various weights and its postures. We measured (1) robotic arm's deflection with various loads from 0 to 80 kg, and (2) variation of carbon bed's deflection with various postures in which the bed is tilted by ±20° roll and pitch angles respectively. Experimental results show that the robotic arm's deflection measured up to 2.6 mm and the variation of carbon bed's deflection measured greater than 0.5 mm when the tilted angle was more than 15°. We, therefore, concluded that the deflection correction function is necessary for our robotic bed especially in the case of headeck treatment in which the tilted angle is needed to be more than 15°.
机译:在本文中,我们将在实验室环境下进行两个初步实验,以确定由于患者体重和姿势的不同,我们的机器人床是否需要进行挠度校正功能。我们测量了(1)在0到80 kg的各种载荷下机械臂的挠度,以及(2)碳床的挠度在各种姿势下的变化,其中,床的倾斜角度分别为±20°侧倾角和俯仰角。实验结果表明,当倾斜角度大于15°时,机械臂的挠度可达2.6 mm,碳床挠度的变化大于0.5 mm。因此,我们得出结论,对于我们的机器人床而言,挠度校正功能是必需的,尤其是在需要倾斜角度大于15°的头部/颈部治疗的情况下。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号