首页> 外文会议>IET International Conference on Biomedical Image and Signal Processing >MRI MOTION ARTIFACT MITIGATION METHODOLOGY USING SPIN ECHO PULSE SEQUENCES ON A 4.7 T SCANNER
【24h】

MRI MOTION ARTIFACT MITIGATION METHODOLOGY USING SPIN ECHO PULSE SEQUENCES ON A 4.7 T SCANNER

机译:MRI运动伪影减缓4.7 T扫描仪中的旋转回波脉冲序列的缓解方法

获取原文

摘要

The movement caused by respiration is one of the most common sources of artifacts in Magnetic Resonance Imaging (MRI). In this work, a methodology to mitigate arti facts related to respiratory movement is proposed using sequences of SE (Spin Echo) pulses in a 4.7 Tesla scanner for small animals without a commercial electronic module for detection of respiratory signals. In this methodology, the RF (Radio Frequency) pulses produced by the scanner in each sequence are detected synchronously with the respiration signal using the respiratory gating technique. To validate the methodology, images of a mechanism built to simulate the respiratory movement are collected and compared to the still images using objective metrics: PSNR (Peak Signal-to-noise-ratio), RMSE (Root Mean Square) and SSIM (Structural Similarity Index Measure). The results showed that the mitigation methodology of the simulated respiratory motion artifacts was efficient and can be used, in future experiments, to mitigate respiratory movement artifacts from abdominal organs of small animals, such as rats and mice.
机译:由呼吸引起的运动是磁共振成像(MRI)中最常见的伪影源之一。在这项工作中,利用4.7特斯拉扫描仪中的SE(自旋回声)脉冲的序列,用于减轻与呼吸运动相关的艺术事实的方法,用于检测呼吸信号的小型动物。在该方法中,使用呼吸门控技术与呼吸信号同步地检测由每个序列中的扫描仪产生的RF(射频)脉冲。为了验证方法,采用模拟呼吸运动的机制的图像,并与使用目标度量的静止图像进行比较:PSNR(峰值信噪比),RMSE(均方根)和SSIM(结构相似度)指数措施)。结果表明,模拟呼吸运动伪影的缓解方法是有效的,可在未来的实验中使用,以减轻小动物的腹部器官的呼吸运动伪影,例如大鼠和小鼠。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号