首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Design and evaluation of a low cost intracranial pressure monitoring system
【24h】

Design and evaluation of a low cost intracranial pressure monitoring system

机译:低成本颅内压监测系统的设计与评估

获取原文

摘要

One of the most life-threatening neural conditions is elevated intracranial pressure (ICP); it is associated with ischemia and poor short- and long-term outcomes. Currently, monitoring systems that accurately measure ICP are either highly invasive or inaccurate. This work explores the design and evaluation of an epidural intracranial pressure monitoring system for low-cost, minimally invasive detection. Our goal is to develop a monitoring system that could also be integrated with an electrocorticography (ECoG) system. To this end we created a minimally invasive epidural ICP monitor for use with a 2 mm burr hole craniotomy. A MEMS piezoresistive sensors is used in the system, and its performance is evaluated for intracranial pressure measurements. Our system is calibrated and tested on the benchtop and demonstrated in vivo using an animal model.
机译:威胁生命的神经疾病之一是颅内压升高(ICP)。它与局部缺血以及短期和长期预后不良有关。当前,准确测量ICP的监视系统是高度侵入性的或不准确的。这项工作探索了用于低成本,微创检测的硬膜外颅内压监测系统的设计和评估。我们的目标是开发一种监控系统,也可以将其与皮层脑电图(ECoG)系统集成在一起。为此,我们创建了一种微创硬膜外ICP监护仪,用于2 mm的钻孔开颅手术。系统中使用了MEMS压阻传感器,并对其性能进行了颅内压测量评估。我们的系统在台式机上进行了校准和测试,并使用动物模型在体内进行了演示。