In this study, Shinnar-Le Roux algorithm is used for designing the radio frequency pulses in MRI in order to reduce the nonlinearity effect of the Bloch equation. Some MRI experiments are conducted to show the effectiveness of Shinnar-Le Roux algorithm by comparing the difference between SLR pulsed slice profile and SINC pulsed slice profile. The experimental results indicates that the slice profile of SLR pulse has smaller ripples and is more uniform than SINC pulse. By reducing the interference of slices, the SLR pulse can improve the imaging quality to a great extent.
机译:在这项研究中,Shinnar-Le Roux算法用于设计MRI中的射频脉冲,以减少Bloch方程的非线性影响。通过比较SLR脉冲切片轮廓和SINC脉冲切片轮廓之间的差异,进行了一些MRI实验以证明Shinnar-Le Roux算法的有效性。实验结果表明,SLR脉冲的条带轮廓比SINC脉冲具有较小的纹波并且更均匀。通过减少切片的干扰,SLR脉冲可以在很大程度上改善成像质量。