首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Estimation of physiological sub-millimeter displacement with CW Doppler radar
【24h】

Estimation of physiological sub-millimeter displacement with CW Doppler radar

机译:连续多普勒雷达估算亚毫米生理位移

获取原文

摘要

Doppler radar physiological sensing has been studied for non-contact detection of vital signs including respiratory and heartbeat rates. This paper presents the first micrometer resolution Wi-Fi band Doppler radar for sub-millimeter physiological displacement measurement. A continuous-wave Doppler radar working at 2.4GHz is used for the measurement. It is intended for estimating small displacements on the body surface resulting from physiological activity. A mechanical mover was used as target, and programmed to conduct sinusoidal motions to simulate pulse motions. Measured displacements were compared with a reference system, which indicates a superior performance in accuracy for having absolute errors less than 10μm, and relative errors below 4%. It indicates the feasibility of highly accurate non-contact monitoring of physiological movements using Doppler radar.
机译:已经研究了多普勒雷达生理感应技术,用于非接触式检测生命体征,包括呼吸频率和心跳频率。本文介绍了首个用于亚毫米生理位移测量的微米分辨率Wi-Fi频带多普勒雷达。测量使用工作在2.4GHz的连续波多普勒雷达。它旨在估计由于生理活动导致的身体表面的微小位移。使用机械动子作为目标,并对其进行编程以进行正弦运动以模拟脉冲运动。将测得的位移与参考系统进行了比较,该系统在绝对误差小于10μm,相对误差小于4%的情况下显示了卓越的精度。这表明了使用多普勒雷达对生理运动进行高精度非接触式监测的可行性。