首页> 外文会议>International Symposium on Communications and Information Technologies >Resolution improvement of point dose distribution in intensity modulated radiation therapy
【24h】

Resolution improvement of point dose distribution in intensity modulated radiation therapy

机译:调强放射治疗中点剂量分布的分辨率提高

获取原文

摘要

In intensity modulated radiation therapy (IMRT), there are verification methods for the point dose distribution, such as two dimensional (2D) detector or film based verification. The 2D detector based verification is known to be easier than the film based one, however the obtained point dose distribution is lower quality. In this paper, we apply a super-resolution method to improve the image quality of point dose distribution obtained from the 2D detector. We conducted an experiment to evaluate the super-resolution images and its ideal reference images. Each dose distribution obtained by proposed method satisfied the criteria in gamma analysis. Although the failed point is slightly observed, clinical problems are probably not caused. Thus the feasibility of the proposed method is confirmed.
机译:在强度调制放射疗法(IMRT)中,存在用于点剂量分布的验证方法,例如二维(2D)检测器或基于胶片的验证。已知基于2D检测器的验证比基于胶片的验证更容易,但是获得的点剂量分布的质量较低。在本文中,我们应用超分辨率方法来改善从2D检测器获得的点剂量分布的图像质量。我们进行了一项实验,以评估超分辨率图像及其理想的参考图像。通过所提出的方法获得的每个剂量分布均满足伽马分析中的标准。尽管可以稍微观察到故障点,但可能不会引起临床问题。因此,证实了所提出方法的可行性。